The ultra-portable Vivo 45 LS life support ventilator is designed to maximize The Vivo 45LS life support ventilator maximizes patient mobility and breathing comfort. It features a built-in humidifier* with heated wire circuit* plus a comprehensive set of modes that includes Auto-EPAP* and High Flow Nasal Therapy* plus SpO2, EtCO2, FiO2 and PtcCO2 monitoring.
* Modification in accordance with FDA’s guidance, Enforcement Policy for Ventilators and Accessories and Other Respiratory Devices During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency.
Settings / Performance
Ventilation Modes
PSV, PSV(TgV), PSV(AE*), PSV(TgV+AE*), PCV, PCV(TgV), PCV(A), PCV(A+TgV), PCV(A+AE*), PCV(A+TgV+AE*), PCV-SIMV, PCV-MPV, VCV, VCV(A), VCV-SIMV, VCV-MPV, CPAP, HFNT*
Patient Modes
Adult, Pediatric
Device Modes
Home, Clinical
Profiles
3 settable profiles
Inspiratory Pressure
4 to 50 cmH2O, max PS of 30 cmH2O above PEEP (Pediatric)
PEEP
2 to 20 cmH2O, IPAP -2 cmH2O or Min IPAP -2 cmH2O
Auto-EPAP (Expiratory Positive Airway Pressure)
2 to 20 cmH2O
Breath Rate/Back-up Rate
4 – 40 bpm (Adult), 6 – 60 bpm (Pediatric)
Inspiratory Time/Backup Insp. Time
0.3 to 5 s (Adult), 0.3 to 2 s (Pediatric)
Rise Time
1 to 9 (PSV & PCV), 50 to 90%, Off (VCV)
Inspiratory Trigger
1 to 9 (PSV & PCV, VCV), Off (PCV, VCV)
Expiratory Trigger
1 to 9
Minimum Inspiratory Time
Off, 0.3 to 3 s (Adult), Off, 0.3 to 2 s (Pediatric)
Maximum Inspiratory Time
0.3 to 5 s, Off (Adult), 0.3 to 2 s, Off (Pediatric)
Tidal/Target Volume
Off, 300 – 2000 ml (Adult), Off, 50 – 400 ml (Pediatric)
High Flow Nasal Therapy*
4 – 60 lpm
Monitoring
Displayed data
Ppeak, PEEP, Pmean, Leakage, MV, Vt, FiO2, % in TgV, Total Rate, Spont Rate, % Spont, SpO2, Pulse Rate, I:E, Inspiratory Time, Rise Time, EtCO2, InspCO2, PtcCO2, Flow, AHI*
Waveforms
Pressure, flow, volume, EtCO2, leakage
Trends over 1, 6, 24, 48 h and 10 days on device screen
SpO2, Ppeak, PEEP, Total Rate, Spontaneous Rate, Leakage, Vte, EtCO2, Flow
Power Supplies
Mains supply
100 to 240 V AC
External DC
19 V ± 6 V
Click-in battery
5.5 hours
Internal battery
2 hours
External Xpac Li-ion battery pack
18 hours
Main alarms
Alarms
High Pressure, Low Pressure, High PEEP, Low PEEP, High Vti, Low Vti, High MVi, Low MVi, High Vte, Low Vte, High MVe, Low MVe, High Breath Rate, Low Breath Rate, Apnea, Disconnection, Rebreathing, Obstruction, High FiO2, Low FiO2, High SpO2, Low SpO2, High EtCO2, Low EtCO2, High InspCO2, High Pulse Rate, Low Pulse Rate, PtcCO2, Power Fail, High Patient Air Temp, Low Patient Air Temp, Low Last Power Source, Crit. Low Last Power Source, Lost Mains Power, Exhalation Valve Control Error, SpO2 Disconnection, SpO2 Signal Lost, Poor SpO2 Signal, CO2 Disconnection, CO2 Accuracy, Check CO2 Adapter, CO2 Sensor, FiO2 Disconnection
Dimensions
W x H x D
8.50” x 6.26” x 5.98” (216 x 159 x 152 mm)
Weight
5.3 lb (2.4 kg)
Noise level (at 10 cmH2O constant pressure)
Less than 30 dB(A)
Connectivity and Software
Connectivity
Ethernet, USB and iLink (plug-in)
Treatment follow up
EveryWare Cloud solution by Breas, PCSW
Plug-and-Play accessories
SpO2, PtcCO2, EtCO2, FiO2, Nurse Call, Remote Alarm, Remote Start/Stop
Other
Humidification*
Integrable water chamber*
Circuits
Disposable leakage circuits with/without heated wires*, disposable exhalation valve circuits, disposable MPV circuits
Intended use: The Vivo 45LS ventilator (with or without the SpO2 and CO2 sensors) is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Specifically, the ventilator is applicable for pediatric through adult patients weighing more than 5 kg (11 lbs.), however, the mouthpiece ventilation modes are for adult patients only. The Vivo 45LS with the SpO2 sensor is intended to measure functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate. The Vivo 45LS with the CO2 sensor is intended to measure CO2 in the inspiratory and expiratory gas. The device is intended to be used in home, institution, hospitals and portable applications such as wheelchairs and gurneys. It may be used for both invasive and non-invasive ventilation. The Vivo 45LS is not intended to be used as an emergency transport or critical care ventilator.
*Modification in accordance with FDA’s guidance, Enforcement Policy for Ventilators and Accessories and Other Respiratory Devices During the Coronavirus Disease 2019 (COVID-19)
Public Health Emergency.Vivo by Breas, Breas, EveryWare by Breas, eSync and the Breas logo are trademarks of Breas Medical AB. © 2022 Breas Medical − All rights reserved. Breas Medical reserves the right to make changes in specifications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your Breas representative for the most current information.
Vivo 45 LS Features
MAR-007760